Fluorescence Fiber Optic Temperature On-line Monitoring Scheme For Medical Industry

Technical Background

The temperature treatment methods in medical industry are microwave therapy, microwave hyperthermia, radiofrequency hyperthermia, and magnetic resonance imaging. In microwave hyperthermia, the main factor affecting the accuracy of temperature measurement results is the interference of microwave or radio frequency. The metal made temperature sensor probe will produce inductive current in microwave field, which makes the measurement results cannot reflect the real temperature of tumor tissues. Whether hyperthermia is effective, the key question is to heat the tumor above the effective treatment temperature. If the tumor temperature is below the treatment temperature, not only has no treatment effect, but also may promote the spread of the tumor, if the temperature is too high, may damage the normal tissue, in about 42℃,1℃ difference can cause multiple changes of cell survival rate. Clinical temperature measurement accuracy within 37℃ ~ 46℃ should be within ±0.2℃, so accurate temperature measurement is the key technology in microwave hyperthermia. 

Comparison of Temperature Measurement Methods

 

Index

 

Fluorescence Fiber Optic Temperature Sensor

 

Conventional Temperature Sensor

 

Thermistor- High Resistance Wire Temperature Sensor

 

Operating environment

 

High voltage under strong electromagnetic field

 

 

High voltage under strong electromagnetic field

 

 

High voltage under strong electromagnetic field

 

 

Materials

 

No metal parts, using fiber optic sensor probe

 

Metal made temperature probe and wire

 

Employ semiconductor thermistors as probes, high resistance wire as signal transmission line, and equipped with a simple measurement circuit

 

Advantages and disadvantages

 

 

1.Anti-electromagnetic interference, high voltage resistance, corrosion resistance

2.Reliable and stable, high precision, anti-explosion and flame-proof

3. Tiny size and light weight

4. Non-toxic

 

1. Under high frequency electromagnetic field, the temperature probe and wire will produce induced current, which increases theirs’ temperature and interferes with temperature.

2. Big error makes it impossible to measure stable temperature.

3.Also bring great harm to microwave equipment

 

1. Poor stability, interchangeability and linearity

2.Poor mechanical strength of high resistance wire, difficult to achieve high accuracy measurement

 

Under the special environment of microwave, radio frequency and high frequency, the fluorescence fiber temperature measurement system can measure the online temperature in real time, which exactly reflects the real temperature of the measured position, and is the optimal temperature measurement method. 

HQ-16/18 Fluorescence Fiber Optic Temperature Monitoring System

Product Description

Name

Performance

Picture

One channel temperature measuring system-

rectum treatment

 

Reading temperature range: 0-80℃Resolution:0.1℃

Temperature accuracy: ±0.3℃

Sensing cable length:< 10m   

Sensing cable diameter:2.3mm

Input power: DC24V

Communication port:485/RS-232Modbus

(protocol / RdLight Logger software)

Analog output:4-20mA or 0-5V

Power consumption:1W

One channel temperature measuring system-Tumor hyperthermia and microwave therapy

 

Reading temperature range: 30-50℃Resolution:0.1℃

Temperature accuracy: ±0.3℃

Sensing cable length:< 10m   

Sensing cable diameter:0.45mm

Input power: DC24V

Communication port:485/RS-232Modbus

(protocol / RdLight Logger software)

Analog output:4-20mA or 0-5V

Power consumption:1W

Six channels temperature measuring system

Reading temperature range: 30-50℃Resolution:0.1℃

Temperature accuracy: ±0.3℃

Sensing cable length:< 10m   

Sensing cable diameter:0.45mm

Input power: 220V

Communication port:485/RS-232Modbus

(protocol / RdLight Logger software)

Analog output:4-20mA or 0-5V

Power consumption:1W

Software Introduction

The supporting software of fluorescence fiber optic temperature measurement products have the functions of reading, storage, drawing and data export, which are convenient for real-time observation and data post-processing.

Temperature Profile

Data Storage Management

 

Scan the QR code to read on your phone

页面修改

页面修改

Time of issue:2020-10-25 00:11:25

Contact Us

 

 

+86-18561133119

 

Address: Yingbin South Street, Lingcheng Economic Development Zone, Dezhou City, Shandong Province

Email:1778780643@qq.com

公众号二维码

Scan wechat

COPYRIGHT©2020 Shandong Qinzhi Electric Co., Ltd.   Back-stage Management 备案号:鲁ICP备2020042284号-1